Product Name and Model
Enteral nutrition pump TS-1001AH (universal pipeline - heating type)
Pumping Method
Disc peristaltic extrusion
Tube Installation Method
Side-mounted nutrient tube for easy use and on-site observation
★ Feeding Mode
Continuous mode, pulse mode, intermittent mode, time mode, scientific feeding mode
★Display Screen
5.0-inch medical capacitive touch screen; combined operation via buttons and touch screen; designed with an angle of 8° for easy observation and operation
★ Infusion rate and volume range
1~2000ml/h, with a minimum step of 1ml/h;
1~9999 ml, minimum step size of 0.1 ml
Counterattack, flushing
Speed: 1~2000 mL/h, minimum step 1 mL/h; Reverse suction and flushing volume: 1~9999 mL, varying by 0.1 mL;
Fast sorting speed range and amount of sorting
100~2000 ml/h, minimum increment 1 ml/h; 1~100 ml
Range of pre-fill speed and pre-fill volume
1 ml/h~2000 ml/h; minimum step size: 1 ml/h; 50 ml
Error range
Speed error: ±5% or 1 ml/h (whichever is greater); Quick drainage speed error: ±10%; Temperature control accuracy: ±1℃;
Time setting range
1 minute to 9 hours, 59 minutes, 59 seconds;
Blocking Sensitivity
40Kpa~100Kpa (300mmHg~750mmHg), four settings adjustable;
★ Heating Range
Range: 30℃ to 42℃, with increments or decrements of 0.1℃
★ Heating Method
Dry heating method, non-water circulation, non-winding heating; involves wrapping the liquid pipeline with flexible heating tubes
★ Storage Records
Stores at least 4000 historical records. The data recorded includes the following information: start time, infusion rate, and infusion mode.
Alarm information and time, total infusion amount, time of infusion cessation
★ Backlight Mode
Auto-backlight mode, Manual backlight mode (adjustable from level 1 to 10)
Backlight delay
Always on, Delayed (1 minute - 10 minutes), Off
Keyboard Lock
The keyboard lock can be turned off or on.
Infusion pressure
Not more than 160 KPa (200 mmHg)
Power consumption
≤60VA includes a heating section
Alarm Function
Battery low warning, battery depleted warning, bubble warning, blockage warning, infusion completion warning, standby timeout warning
Electrical alarm, pipeline error alarm, over-temperature alarm, low temperature alarm, abnormal speed alarm, door open alarm, pipeline switching error alarm, hard
Fault alert
Time Setting
Date and time can be adjusted to facilitate effective historical record retrieval
External Interface
1 USB port, 1 external power supply port, 1 heating device aviation socket port
Appearance dimensions and weight
127 mm * 145 mm * 184 mm; 1.6 kg
Safety classification / Waterproof rating
Class I, internal power supply equipment with anti-defibrillation CF application section; IPX3
Battery specifications
DC1 11.1V, 2,600mAh; after charging, it can operate continuously at a flow rate of 25 mL/h for more than 6 hours
★Expiry Date
Validity period: 10 years
Work Environment
Environmental temperature: 5℃~40℃; Relative humidity: 20%~90%; Atmospheric pressure: 70.0kPa~106.0kPa
Power supply: AC input ranging from 100~240V
50Hz/60Hz
Storage and transportation conditions
Relative humidity ≤ 93%, temperature: -20℃ to +55℃; atmospheric pressure range: 50.0 Kpa to 106.0 Kpa;
Other Features
The engineering mode features password protection.