Packaging information
CTN Size:
1000
×
58
×
87
cm
G.W./CTN:68 kg
QTY/CTN:1 pcs/carton
● Servo Control Mode: air mode;
● Instant ParameterDisplaying: set temp, air temp, heat power rate;
● Audible & Visual Alarms:power failure, sensor failure, higher/lower temp,
air temperature override, fan failure;
● Power failure reset memory and self-test function;
● Hood with dual slope: 4hand-ports;
● Optional skin temperature display;
● Optional humidity display.
Specifications of hood
Mattress to Hood: 450 mm
Observation side of hood: Two incline
Operate port: 4 (2 Iris ports)
Tubing grommets: 2
Front access panel: 826×310 mm
Mattress Tray Size: 630×360 mm
Technical Specifications
Power requirement ≤400VA
Temperature control mode: Air servo control
Air Temp Control Range: 20.0℃-37.0℃
Temperature fluctuation: ≤0.5℃
Mattress temperature uniformity: ≤0.8℃
Temperature rise time: ≤45 min
Noise level within hood environment: ≤55dB
Alarms
Power failure alarm
Fan failure alarm
Sensor failure alarm
High air temperature alarm: +3.0℃
Low air temperature alarm: -3.0℃
Over air temperature alarm: ≤38.0℃
Operational condition
Environment temperature: +20℃-+30℃
Environment relative humidity: 30%℃-75%
Environment air velocity of flow: <0.3m/s
Others
Mattress tray Max. load 5kg
Transfusion pole Max. load 2kg
Update time:20251223175556