1. Specifications and Parameters:
Auto Shutdown: 3-minute auto shutdown (can be set)
Operating temperature: 0-40°C
LCD display: 5-digit LCD display
Backlight: White Display (Optional)
Power supply: 2X AAA batteries / external DC 6V power supply
Maximum Display Box Size: 166X142X81MM
Display Box Size: 166X126X32MM
Tray: 113X99MM
Single unit weight: 225G
Tare Function
Unit Selection: (UNIT): KG (Kilograms)/G (Gram)/LB (Pounds)/CA (Cantons) T (Tai pounds)
Automatic Low Battery Alarm: Waterproof Touch Button Design
2. Operating Instructions:
Insert the 2X "AAA" batteries into the battery compartment following the instructions, or connect an external power source to the mains (DC=6V). Place the scale on a stable, wind-free surface.
1. Press the "ON" key, and wait for 5 seconds to display 0.0.
2. Place the heavy object on the tray surface, and when the S symbol appears in the upper left corner, it indicates the accurate weight of the object being weighed;
3. Press the UNIT key to change the measurement unit, G/KG/LB/CA(T)
4. Press TARE to force zero.
5. Press to turn on/off the power when the device is in standby mode.
6. Calibration function:
Press the ON button for 3 seconds until the LCD displays "0" and then release the button. Press the tare key to confirm and display a flashing weight value (e.g., 3000.0). At this point, place a 3000g weight on the tray. (If you do not have an appropriate weight, you can select the appropriate weight value by pressing the TARE key for calibration.) When the value stops flashing and displays the current weight value, the calibration is complete. (If the temperature difference is too large or the weighing is inaccurate for other reasons, use this function.)
Section 3: Unit Conversion
1KG = 1000G = 1.6534CA = 2.2LB = 1.6667台斤
1CA港斤 = 16港两 = 604.78G
1 jin = 16 taels = 600 grams
"1LB(磅)=453.59G=16OZ
Four. Points to Note
1. Place the scale on a level and stable surface in a vibration-free and wind-free environment.
2. Please do not place any object weighing more than the rated weight on the pallet, as this may damage the sensor. When placing objects, please do not impact the weighing pallet.
3. When the display shows the battery symbol, please replace the battery immediately.
4. Please use this product in an environment with a temperature difference of 0-40°C, as a large temperature difference may cause measurement errors.
If there is a large temperature difference, please use a standard weight to calibrate this product. When calibrating, ensure that this product is in the same environment for at least 30 minutes.
5. When the weight of the pallet exceeds the rated weight, the "FULL" symbol will appear.
6. This product is designed with touch buttons. After continuous rapid button pressing, it may experience a deadlock (key lock) phenomenon. Please turn off the power for 1 minute and then turn it back on to resolve the issue.